BTX Gemini X2 Twin Wave Elektroporator
Square wave pulse ranges
Voltage range: LV mode 5 to 500/1 V res, HV mode 505 to 3000/5 V res
Voltage accuracy: LV mode 5%, HV mode 5%
Pulse length range: LV mode 10 µs to 999 µs/1 µs res, 1 ms to 999 ms/1 ms res, HV mode 10 µs to 600 µs/1 µs
Multiple pulses: LV mode 1 to 120 (10 per sample), HV mode 1 to 36 (3 per sample), pulse interval 0.1 s to 10 sec
Capacitance in µF: LV mode 3775 µF, HV mode 85 µF, voltage drop 20% load < 1 K Ω, 10% load > 1 K
Exponential decay pulse ranges
Voltage range: LV mode 5 to 500/1 V res, HV mode 505 to 3000/5 V res
Voltage accuracy: LV mode 5%, HV mode 5%
Time constant range: LV mode 1 ms to 5,158 S, HV mode 0.5 ms to 133,875 ms
Multiple pulses: Multiple pulses 1 to 12 (Rint <100 ohm), 1 to 24 (Rint >100 ohm), pulse interval 5 to 30 sec
Capacity in µF: LV mode 25 to 3275 µF/25 µF res, HV mode 10, 25, 35, 50, 60, 75, 85 µF
Internal resistance (parallel to load): LV 25 to 1575 Ω / 25 Ω res, HV 50 to 1575 Ω / 25 Ω res
Sample resistance (load) Exponential decay or square: LV mode PL ≤ 100 ms, load ≥ 8-9 Ω, PL > 100 ms, load ≥ 100 Ω, HV mode load ≥ 40 Ω
Charging time: LV < 7 sec, HV < 4 sec
Monitoring: pulse voltage, width, droop% and sample resistance
Programmability: Storage of more than 1000 protocols
Arc control: Yes
Interlock switches: Yes
Pre-pulse sample resistance control: Yes
Pulse overcurrent protection: Yes
PC control: Yes
PC communication: Yes
Logging report: Yes
Remote control: Yes
Foot switch operation: Yes
Dimensions (HxWxD): 324 x 286 x 216 mm
Weight: 7.3 kg