Agilent 7500CE ICP-MS
Agilent 7500 Series ICP-MS systems use efficient low-flow nebulisers and Peltier-cooled Scott spray chambers. All systems support the optional Integrated Sample Introduction System (ISIS) for added flexibility and productivity.
Compact tabletop model enables easy integration into existing laboratory spaces
Full stainless steel chassis for superior corrosion resistance
Open architecture sample introduction area simplifies interfacing with alternative sampling accessories
Peltier-cooled spray chamber for optimal stability and minimal oxide interference
Fully automatic, computer-controlled torch positioning in all three axes
ShieldTorch System (STS) for highest sensitivity, best ion energy control
Agilent active mass flow controllers (AMFC) for precise control of all plasma and collision/reaction cell gases
High efficiency digital drive 27 MHz ICP RF generator is fully solid state and maintenance free
True hyperbolic profile quadrupole rods for excellent ion transmission and superior sensitivity to abundance
Highest frequency, digitally synthesised 3 MHz RF Quadrupole Driver for enhanced mass spectrometer performance
Simultaneous dual-mode detector with exclusive high-speed amplifier provides true nine orders of linear dynamic range
Nebuliser: MicroMist glass concentric nebuliser
Spray chamber: Quartz Scott-type
Peristaltic pump: Agilent design with high precision, 10-roll, 3-channel pump for precise computer-controlled pumping of sample, internal stand and spray chamber drain. The pump is located next to the spray chamber for the fastest flush in/out and minimum pipe volume.
Cone: Nickel
Extraction lens: Double on-axis extraction lens outside high vacuum area for easy maintenance allows conventional and gentle extraction mode.
Mass range: 2-260 AMU
RF frequency: 3 MHz
Minimum waiting time: 100 μs
Overload sensitivity: low mass <5 X 10-7, high mass <1 X 10-7
Detector electronics: Agilent's unique log amplifier circuitry has a high-speed analogue mode for transient signals and a dynamic range of nine orders. Minimum latency: 100 μs (in both pulse counter mode and analogue mode) Working concentration range: detection limit up to 500 ppm Maximum measurable concentration can be extended to> 1000 ppm in He mode (7500ce, 7500cs) - without adjusting the resolution of the analyser.
System includes:
- Agilent 7500 ICP-MS
- Cetac ASX-5200 Autosampler
- Agilent G1879 Heat Exchanger
- Edwards E2M18 Vacuum Pump
- PC
- Software
- Manuals